SCPI Command Description
:ABORt Subsystem
The :ABORt command is a single command subsystem. There are no subcommands or
associated data parameters, as shown below. The :ABORt command, along with the
:TRIGger and :INITiate commands, comprise the Trigger group of commands.
Command Parameters
Unit
(default)
Remark
:ABORt
:ABORt
:ABORt
This command causes the List or Step sweep in progress to abort. Even if
INIT:CONT[:ALL] is set to ON, the sweep will not immediately re-initiate.
:DISPlay Subsystem
The :DISPlay subsystem controls the display of the front panel data display.
Command Parameters
Unit
(default)
Remark
:DISPlay[:WINDow]:TEXT:[STATe]
ON|OFF|1|0 FW2.4
:DISPlay:REMote
ON|OFF|1|0 FW2.4
:DISPlay:WINDow:TEST ON|OFF|1|0
:DISPlay[:WINDow]:TEXT:[STATe] ON|OFF|1|0
:DISPlay[:WINDow]:TEXT:STATe?
Turns on/off the parameter display of the BNC MODEL 835-3 front panel data display.
:DISPlay:REMote ON|OFF|1|0
:DISPlay:REMote ON|OFF|1|0
:DISPlay:REMote?
Turns on/off the display update of the BNC MODEL 835-3 front panel data display.
:INITiate Subsystem
The :INITiate subsystem controls the state of the BNC MODEL 835-3 trigger system. The
subsystem commands and parameters are described below. The :INITiate commands,
along with the :ABORt and :TRIGger commands, comprise the Trigger Group of
commands.
Command Parameters Unit
(default)
Remark
:INITiate[:IMMediate]
ON
:INITiate:CONTinuous
ON|OFF|1|0 OFF
:INITiate[:IMMediate]
Sets BNC MODEL 835-3 trigger to the armed state.
:INITiate:CONTinuous ON|OFF|1|0