Bnc 1201 - 6 1/2 Digit Digital Multimeter Manual do Utilizador Página 142

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 177
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 141
143
resolution. The reading is sent to the output buffer. For frequency
measurements, the meter uses only one “range” for all inputs between 3Hz and
300kHz. With no input signal applied, frequency measurements return “0”.
MEASure:PERiod? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a period measurement with the specified range and resolution.
The reading is sent to the output buffer. For period measurements, the meter
uses only one “range” for all inputs between 0.33 seconds and 3.3 μsec. With
no input signal applied, period measurements return “0”.
MEASure:CONTinuity?
Preset and make a continuity measurement. The reading is sent to the output
buffer. The range and resolution are fixed at 1kΩ and 5 ½ digits respectively.
MEASure:DIODe?
Preset and make a diode measurement. The reading is sent to the output buffer.
The range and resolution are fixed at 1Vdc with 1mA current source and 5 ½
digits respectively.
MEASure:TEMPerature?
Preset and make a temperature measurement with RTD. The reading is sent to
the output buffer.
MEASure:TCOuple?
Preset and make a temperature measurement with thermocouple. The reading
is sent to the output buffer.
C.4 The CONFigure Command
CONFigure:VOLTage:DC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for DC voltage measurements with the
specified range and resolution. This command does not initiate the
measurement.
CONFigure:VOLTage:DC:RATio {<range>|MIN|MAX|DEF },{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for DC:DC ratio measurements with the
specified range and resolution. This command does not initiate the
Vista de página 141
1 2 ... 137 138 139 140 141 142 143 144 145 146 147 ... 176 177

Comentários a estes Manuais

Sem comentários