18 SAM 935™ Instruction Manual
2. Select
Area Monitor Setup
. The Setup menu takes up several screens; the arrow keys will scroll
through the setup screens:
3. Use the arrow keys to highlight
Select Dose Rate Units
and press ENTER.
4. Select the units of measurement you want to use when monitoring: F2 for REMs per hour (R/hr) or
F3 for Sievert per hour (Sv/hr). The message line above the function key labels tells you the cur-
rently selected units.
5. The display will immediately return to the Monitor Setup menu. Select
Select Dose Rate Trigger
from the menu:
Select
Dose Rate Units
to choose the units of
measurement.