SECTION 2
OPERATING INFORMATION
General
POWER UP
When power is first applied, the mainframe performs a test to determine if a module is installed. If a
module is present, the limits for each module dependent parameter are transferred to the mainframe.
The memory is checked for any errors and the parameters in memory 0 are loaded into the mainframe.
If a module is not present, the parameters are set as indicated in Table 2-1.
Table 1-4. 6040 Mainframe Default Settings
LCD Power On Sequence
When the 6040 is first turned on, the display will show the software version number. If a module is
present, the LCD will then momentarily display "MEMORY CHECK," followed by the module I.D.
display. Without a module, the display will read "6040 No Plug-In."
MODE:
Pulse
TRIG: Single Cycle (with other values set as follows)
Internal Trigger Rate = 1 kHz
External Trigger Threshold = 0 V
External Trigger Slope: +
TIMING: Width = 1 µs
Delay = 1 µs
Single Pulse
GPIB/RS-232: IEEE-488 Address = 6
Baud Rate = 1200
Full Duplex
Remote Enabled