:CALC2:LIM:FCO
:CALC2:LIM:PCO
:CALC3:…
Note
The re-calculation is done automatically when any change is made to any of
the :CALCulate[1|2] commands when :CALC:IMM:AUTO is ON.
:CALCulate2:IMMediate:AUTO <Boolean>
Sets if the re-calculation will automatically be done whenever there’s any
change made to any of the :CALCulate[1|2] commands.
When it is set to OFF, the :CALCulate[1|2] command produces new results when
the command is executed, even when new data is not acquired. And this allows
the user to have the new results on the same current data right after new
configuration changes.
Return
An ASCII-encoded byte. 0 means OFF, and 1 means ON.
Notes
This will be set to OFF after the counter has been turned off or a remote interface
reset.
Affects all the recalculation command settings.
The only limit result that can be post-processed is :CALC2:LIM:FAIL?. The limit
counts (:CALC2:LIM:FCO and :CALC:LIM:PCO) reflect measurements that were
limit-tested at time of data acquisition.
:CALCulate2:LIMit:CLEar:AUTO <Boolean>
Sets if the limit test results to be cleared when an :INITiate[:IMMediate] or
an :INITiate:CONTiunuous ON command is encountered.
When it is set to OFF, only will the :CALC2:LIM:CLE[:IMM] command clear the limit
testing results.
Return
A single ASCII-encoded byte. 0 means OFF and 1 means ON.
Notes